Cell Ranger3.1, printed on 11/05/2024
Though cellranger mkfastq is the preferred option for converting BCLs to Cell Ranger--compatible FASTQs, there are a few cases where you may need to use Illumina's bcl2fastq directly to generate FASTQs. This is necessary if:
You may also choose this method if bcl2fastq is more tightly integrated into your sequencing workflow.
Demultiplexing Chromium data with Illumina bcl2fastq requires the correct specification of the sample sheet and command-line options. This guide walks you through what you'll need to do to generate Cell Ranger-compatible FASTQs.
You need to create a sample sheet in order to get bcl2fastq to correctly embed the names of samples into output FASTQ files. There is a key difference to keep in mind when creating sample sheets for a Chromium run. Each Chromium sample index set is actually a blend of four different sequence oligos, and each oligo must be represented as a separate row in the sample sheet. This means that for every sample being demultiplexed from the flowcell, there should be four lines in the sample sheet.
The tool below helps you accurately generate data lines for your sample sheet. When you plan an experiment, you should know the name of the sample index set used for each sample, which comes from the reagent kit (such as "SI-P01-A2"). For each sample, enter its lane, sample name, and sample index set below, and then press 'Add'. When you're done, you can either copy and paste comma-separated output directly into a text editor to create a sample sheet CSV, or copy/paste tab-separated output into a spreadsheet such as Microsoft Excel:
If you are just running a single sample in a lane, then you can have a single line with the index blank, though bcl2fastq will include reads associated with any sample index.
Illumina bcl2fastq must be called with the correct --use-bases-mask
argument and other arguments in order to properly demultiplex and output FASTQs for all the reads in a Chromium library.
In the examples below, ${FLOWCELL_DIR}
is the directory that contains a flowcell's Data folder, ${OUTPUT_DIR}
is the directory that you want to output FASTQs to, and
${SAMPLE_SHEET_PATH}
is the path to the sample sheet CSV you created.
bcl2fastq Version 2.17 or higher
This is the most common case, for sequencers running RTA 1.18.54 and higher.
$ bcl2fastq --use-bases-mask=Y150,I8,Y150 \ --create-fastq-for-index-reads \ --minimum-trimmed-read-length=8 \ --mask-short-adapter-reads=8 \ --ignore-missing-positions \ --ignore-missing-controls \ --ignore-missing-filter \ --ignore-missing-bcls \ -r 6 -w 6 \ -R ${FLOWCELL_DIR} \ --output-dir=${OUTPUT_DIR} \ --interop-dir=${INTEROP_DIR} \ --sample-sheet=${SAMPLE_SHEET_PATH}
Version 1.8.4
$ configureBclToFastq.pl --use-bases-mask=Y150,I8,Y150 \ --fastq-cluster-count=20000000 \ --no-eamss \ --ignore-missing-bcl \ --ignore-missing-control \ --ignore-missing-stats \ --mismatches=1 \ --input-dir=${FLOWCELL_DIR}/Data/Intensities/BaseCalls \ --output-dir=${OUTPUT_DIR}
In both cases, if you want to limit bcl2fastq to a subset of lanes, you need to supply values
to the --tiles
argument.
If you add extra bases to a sample index read, you need to account for this in the
--use-bases-mask
argument. For example, if you ran a sample index read with 9 bases, you need to
truncate the last base in order for Cell Ranger to run correctly.
You can exclude a single base by adding a single n
character to
the read argument, or adding n*
to exclude all bases after a certain position. See below:
Single Cell 5′ chemistry
Read | Desired | Actual | Argument |
---|---|---|---|
i7 Index Read (I1) | 8 | 9 | I8n |
After generating FASTQs, you should be able to follow the pipeline instructions, with one caveat. Instead of using
the --indices
argument to cellranger to select samples, you will use the --sample
argument. The
value of --sample
should be the name of the sample, which should have been in the Sample_Name
column in your sample sheet. The value of --fastqs
should be ${OUTPUT_DIR}/${PROJECT_NAME}
where ${OUTPUT_DIR}
is as defined above
and ${PROJECT_NAME}
is the value in the Sample_Project
column in the sample sheet.